GUI Reference
R&S
®
ZNB/ZNBT
400
User Manual 1173.9163.02 ─ 38
Figure 5-33: Scalar Power Calibration
3. Perform a complementary source flatness calibration (optional).
Figure 5-34: Source Flatness Calibration (LO: external generator)
During source flatness calibration the DUT should be connected with all ports
defined in the mixer setup. First the LO ports are calibrated. In the second step the
source flatness of the RF and IF ports is calibrated while the LO ports are activa-
ted.
5.6 Trigger Softtool
The "Trigger" softtool provides trigger and sweep control settings.
Access:
CHANNEL – TRIGGER hardkey
5.6.1 Trigger Tab
Selects the source of the trigger signal and provides additional trigger settings.
Trigger system of the analyzer
The trigger system is used to synchronize the analyzer's actions with events that can
be provided by an internal or external signal or user-generated ("Manual Trigger").
Triggered measurements are an alternative to the default mode ("FreeRun", "Continu-
Trigger Softtool