[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>, <TestPort2>...
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are
auto-detected
at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg>
Number of the port assignment
Parameters:
<TestPort1>
First test port number
<TestPort2>
Second test port number ...
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See
Chapter 4.5.5.4, "Multiple Port Assignments"
on page 164 for back-
ground information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg>
Number of the port assignment
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
"Start Cal Sweep / Abort Sweep"
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with
, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
)
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO: