GUI Reference
R&S
®
ZNB/ZNBT
607
User Manual 1173.9163.02 ─ 38
Independent of the state of the "Show Cal Kit Label" flag, cal kit labels are displayed in
several other manual calibration dialogs.
Delete Cal Pool / Delete All Cal Kits
Deletes all calibration data and all cal kit data. See
Chapter 5.4.4.3, "Calibration Man-
Remote command:
n/a
Search Path for additional Cal Kits and Connector Types
Contains the name and path of a special directory for cal kit files (
*.calkit
). All cal
kit files in the special directory will be (re-)loaded automatically as predefined kits (i.e.
read-only kits which cannot be modified) every time the VNA application is started. It is
possible to select the default cal kit directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration
or any
other directory. "None" means that no cal kit files are loaded on start-up.
Use the special directory to make sure that you do not have to import kits manually,
even after terminating the VNA application improperly, in which case previously impor-
ted cal kit files will not be stored to the recall set file.
Remote command:
User Interface tab
Provides general user interface configurations.
Sounds / Transparent Info Fields / Show Sweep Symbols
The buttons switch the instrument messages, acoustic messages, transparent info
fields for markers and trace statistics, and sweep symbols on or off.
●
Sounds are generated when the analyzer generates a notice/status message or a
warning (alarm sounds) or during calibration.
●
Transparent info fields do not hide an underlying trace.
●
The sweep symbols are arrows pointing downward onto the trace; they are dis-
played if the sweep time exceeds a lower limit (e.g. for a large number of points or
a small measurement bandwidth).
Setup Softtool