GUI Reference
R&S
®
ZNB/ZNBT
355
User Manual 1173.9163.02 ─ 38
Start Meas
Starts the voltage and current measurement.
Note that "Start Meas" does not automatically activate the "Output Voltage" and
"Range" / "Shunt" settings of the current channel. Use
to activate them manually.
The measurement (=sampling) time can only be set via remote command. It is not
channel-specific and applies to RFFE and GPIO measurements.
Remote command:
CONTrol<Ch>:RFFE:TEST:SENSe:TRIGger
CONTrol<Ch>:RFFE:TEST:TIME
Voltage, Current
Results of the voltage/current measurements on the related RFFE pins.
Remote command:
5.5.1.6
GPIO Configuration
Access:
●
of the "RFFE Config" dialog if the analyzer is equipped with a R&S
ZN-
B15/-Z15 Var. 02
●
of the "RFFE Config" dialog if the analyzer is equipped with a R&S
ZN-
B15/-Z15 Var. 03
Basic GPIO Configuration
The (left part of the) GPIO config table gives access to the channel-specific GPIO volt-
age settings.
Figure 5-25: GPIO Configuration for HW Var. 02
Channel Config Softtool