CONTrol:HANDler:SWEepend
<SweepEnd>
Specifies the event that will cause the /SWEEP END line (pin 34) to go low; see
ter 10.2.4.4, "Timing of Control Signals"
Parameters:
<SweepEnd>
SWEep | CHANnel | GLOBal
SWEep
–
every time a sweep is complete
CHANnel
–
when all the sweeps for each channel are complete
GLOBal
–
when all sweeps in all channels are complete
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
*RST:
n/a (default: GLOBal)
Example:
See
CONTrol<Ch>:RFFE:TEST:OUTPut
Applies the current output voltage and current range (shunt) settings to the RFFE/
GPIO pins.
CONTrol<Ch>:GPIO:VOLTage:OUTPut
.
Suffix:
<Ch>
.
Channel number
Usage:
Event
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:RFFE:TEST:TIME
<MeasTime>
Sets the measurement time for the voltage/current measurements on the RFFE and
GPIO pins.
Same functionality as
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted: the measurement time
is valid for all channels.
Parameters:
<MeasTime>
Measurement (= sampling = averaging) time
Range:
95 μs to 100 ms
*RST:
100 ms
Default unit: s
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See