Command Reference
R&S
®
ZNB/ZNBT
901
User Manual 1173.9163.02 ─ 38
Example:
*RST; DISP:MENU:KEY:EXECute ':Cal:Calibration:
StartCalUnit'
Activates the
(provided that a cal unit is
connected).
Usage:
Setting only
Tip:
When working with the
, switch to raw mode ("Options > Raw
mode" in the IECWIN32 GUI) before executing this command.
DISPlay:MENU:KEY:SELect
<TabID>
Activates the softtool tab with identifier <TabID>.
Use
DISPlay:MENU:KEY:TOOL:CATalog?
to display the available identifiers.
Note
that no error is generated if a valid <TabID> is specified but the tab cannot be
activated for any other reason.
Setting parameters:
<TabID>
Identifier of a softtool tab
Example:
*RST; DISP:MENU:KEY:SEL ':Meas:SParams'
Activates the "S-Params" tab of the "Meas" softtool.
Usage:
Setting only
Tip:
When working with the
, switch to raw mode ("Options > Raw
mode" in the IECWIN32 GUI) before executing this command.
DISPlay:MENU:KEY:TOOL:CATalog?
Displays the identifiers of the available softtool tabs as a comma-separated list of
strings.
Use
Example:
DISPlay:MENU:KEY:TOOL:CATalog?
returns the list of identifiers, from
':Applic:ExternalTools'
to
':Trigger:SweepControl'
Usage:
Query only
DISPlay:RFSize
<RelFontSize>
Defines the size of the fonts in the diagram on a relative scale.
Parameters:
<RelFontSize>
Relative font size
Range:
80 % to 170 %
*RST:
100 %
Default unit: percent
SCPI Command Reference