Command Reference
R&S
®
ZNB/ZNBT
1062
User Manual 1173.9163.02 ─ 38
<Seg>
Sweep segment number
Parameters:
<ResBandwidth>
Resolution bandwidth
Range:
See above
Increment: 1-1.5-2-3-5-7 steps
*RST:
10 kHz
Default unit: Hz
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
<Boolean>
Selects common or independent "Meas. Bandwidth" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent bandwidth settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
OFF
–
reset the bandwidth in all sweep segments to the band-
width for unsegmented sweeps, defined via
. ON will not restore the previous val-
ues.
The parameter is automatically switched to ON when a band-
width is entered using
*RST:
OFF
Example:
*RST; :SENS:SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus 10 kHz measurement bandwidth.
SEGM:BWID 1 MHZ
Increase the resolution bandwidth to 1 MHz.
SEGM:BWID:CONT OFF
Couple the bandwidths in all segments and reset the bandwidth
in segment no. 1 to the initial value.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect
<Selectivity>
Defines the "Selectivity" of the IF filter used in sweep segment no.
<Seg>
. At the same
time, the command activates individual selectivity settings for all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
SCPI Command Reference