GUI Reference
R&S
®
ZNB/ZNBT
312
User Manual 1173.9163.02 ─ 38
The upper part of the calibration screen presents
●
calibration sweep diagrams for the currently measured S-parameter during stand-
ards measurement,
●
power trace diagrams during the power calibration sweep
The lower part displays the calibrated port (pairs) with their related measurements and
visualizes the measurement progress.
Calibration Sweep Diagrams
During the calibration sweep each diagram contains a single S-parameter trace and a
typical result trace for the measured calibration standard.
If switch matrices are involved, a sweep is performed for every possible signal path
and for each of these paths a separate S-parameter trace is shown (see
ter 4.7.20.5, "Multiple Paths and Calibration"
Figure 5-22: Multiple traces due to multiple paths
Cal Softtool