Concepts and Features
R&S
®
ZNB/ZNBT
146
User Manual 1173.9163.02 ─ 38
Figure 4-5: Adapter Removal vs. UOSM
The obtained adapter characteristics are mathematically removed from the obtained
error coefficients. Uncertainties arising from a non-ideal characterization of the
unknown through almost cancel, whereas they add up in the UOSM technique. As a
consequence, Adapter Removal will provide more accurate results.
This holds up to a transition point where uncertainties due to connectivity issues
increase.
Calibration