GUI Reference
R&S
®
ZNB/ZNBT
376
User Manual 1173.9163.02 ─ 38
5.5.3.2
AGC Manual Configuration Dialog
The "AGC Manual Configuration" dialog allows to configure the Automatic Gain Control
for individual sweep segments, drive ports and receivers. This button is enabled in
"Manual"
only.
Access:
CHANNEL – CHANNEL CONFIG > "Mode" > "AGC Mode" – "Manual Con-
fig..."
The manual AGC settings are persisted and reused for subsequent measurement
sweeps, which may increase measurement speed compared to "Auto" mode while not
compromising measurement quality.
●
Without the
option only the AGC of the measurement
receivers (the b-waves) can be statically set to "Low Dist" or "Low Noise"; the refer-
ence receiver AGC mode is always set to "Auto" (see
With the option available, also the AGC of the reference receivers (the a-waves)
can be set statically.
●
The increase in measurement speed for settings "Low Dist" and "Low Noise" is not
achieved if the AGC mode of one of the receivers is set to "Auto".
Range
If
is enabled, each sweep segment can be configured separately.
Remote command:
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:ALL
Drive-port specific settings
Drive Port
,
a
,
b
j
: Selects the AGC mode for the respective drive port, a and b wave and
receiver.
Auto (column)
: Enables the automatic mode for the corresponding drive port, disa-
bling the manual configuration for the related a- and b-waves.
Channel Config Softtool