GUI Reference
R&S
®
ZNB/ZNBT
253
User Manual 1173.9163.02 ─ 38
●
"Selectivity" defines the selectivity of the IF filter used for each sweep segment.
See
●
"LO < > RF" defines whether the analyzer measures the segment with a local oscil-
lator frequency LO below or above the RF input frequency, or if it should use auto-
matic "spur avoidance". This overrides the channel's "Image Suppr." setting for par-
ticular segments; see
●
"Segment Bits" enables the definition of a segment-dependent four-bit binary value
to control four independent output signals at the USER PORT connector (lines 16,
17, 18, 19; see
on page 1252). The output sig-
nals are 3.3 V TTL signals which can be used to differentiate between up to 16
independent analyzer states.
Setting the segment bits does not change the analyzer state.
●
"Time" defines the sweep time for each segment. The default configuration for a
new segment is equal to the sweep time setting for unsegmented sweeps; see
When "Time" is checked two new columns appear in the table. The first column
reads "Segment Time" or "Meas Delay", depending on the selected radio button
below the "Time" checkbox. The second column reads "Auto" and is used to acti-
vate automatic sweep time setting.
"Segment Time" is the total measurement time for the sweep segment. The mini-
mum segment sweep time to be set is equal to the estimated value in "Auto" mode.
"Meas Delay" sets a delay time allowing the DUT to settle before the hardware set-
tings of the analyzer are changed and a new partial measurement is started.
Changing the "Meas Delay" modifies the "Segment Time" and vice versa.
"Auto" minimizes the sweep time. If this option is checked, the columns "Segment
Time" or "Meas Delay" cannot be edited. "Segment Time" indicates the estimated
sweep time, depending on the current measurement settings, the "Meas Delay" is
0 s. The segment sweep time and point delay values are maintained until changed
explicitly if "Auto" is switched off.
Remote command:
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:SWEep:GENeration
[SENSe<Ch>:]SWEep:GENeration:ANALog:CONDition?
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:DEFine
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect
[SENSe<Ch>:]SEGMent<Seg>:INSert
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?
CONTrol:AUXiliary:C[:DATA]
Sweep Softtool