Command Reference
R&S
®
ZNB/ZNBT
1142
User Manual 1173.9163.02 ─ 38
Note:
since firmware V2.80 this command raises an error if no GPIB interface (B10
option) is avalable.
Parameters:
<Terminator>
LFEoi | EOI
LF character sequence with or without EOI recognized.
Example:
SYST:COMM:GPIB:RTER EOI
Set the terminator to EOI.
SYSTem:COMMunicate:NET:HOSTname
<Hostname>
Sets or gets the host name of the instrument.
Parameters:
<Hostname>
Host name
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
<Address>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...
). This command is not necessary if
only one cal unit is connected.
Parameters:
<Address>
Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Queries the names (USB addresses) of all connected calibration units.
Example:
See
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Usage:
Query only
Manual operation:
See
SCPI Command Reference