Command Reference
R&S
®
ZNB/ZNBT
774
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage:
Event
Manual operation:
See
CALCulate<Chn>:RIPPle:CONTrol:DOMain
<SweepType>
Deletes the existing ripple limit ranges and (re-)defines the physical units of the stimu-
lus values of the ripple limit lines. The unit of the ripple limit is defined via
CALCulate<Chn>:RIPPle:RDOMain:FORMat
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<SweepType>
FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG | PLIN | PLOG |
PSINgle
Keywords for the units of the stimulus values; frequency, power,
and time units. The selected unit must be compatible with the
sweep type (
erwise the ripple limit lines cannot be displayed and no ripple
limit check is possible.
*RST:
FLIN
Default unit: Hz (for FLIN, FLOG, FSEG, and FSINgle), s (for
TLIN and TLOG), dBm (for PLIN, PLOG and PSIN-
gle).
Example:
SWE:TYPE POW
Select a power sweep.
CALC:RIPP:CONT:DOM PLIN
Delete all existing ripple limit ranges and select level units for the
ripple limit of the active trace.
CALC:RIPP:DATA 1, -10, -5, 3
Define and enable a ripple limit range in the stimulus range
between -10 dBm and -5 dBm, assigning a ripple limit of 3 dB.
Usage:
Setting only
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DATA
<RippleLimRange>...
Adds and enables/disables an arbitrary number of ripple limit ranges, assigning the
stimulus values and the ripple limits. See
Chapter 4.4.1.2, "Rules for Ripple Test Defi-
Note:
This command does not overwrite existing ripple limit ranges. The defined
ranges are appended to the range list as new ranges. Use the
CALCulate<Chn>:RIPPle:SEGMent<Seg>...
commands to change existing ripple
limits.
SCPI Command Reference