Concepts and Features
R&S
®
ZNB/ZNBT
224
User Manual 1173.9163.02 ─ 38
●
Power calibration: An external generator can provide the reference signal for a
source or receiver calibration. A typical example is a receiver power calibration
using a measured wave b
1
.
External generators must be configured with their connection type and device address
before they are available as additional sources "SYSTEM SETUP > External Devices >
Generators". Configured generators appear in many control elements of the R&S
ZNB/