Command Reference
R&S
®
ZNB/ZNBT
807
User Manual 1173.9163.02 ─ 38
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G<Cmp> in the different circuit models for balanced
port deembedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Number of conductance in circuit model. The total number of
conductances depends on the selected circuit model.
Parameters:
<Conductance>
Conductance G<Cmp> for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp>
<CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp>?
<CircuitModel>
Specifies the inductance value L<Cmp> in the different circuit models for balanced port
deembedding.
In the query form, the <Inductance> parameter must be omitted. The command returns
the inductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Number of inductance in circuit model. The total number of
inductances depends on the selected circuit model.
SCPI Command Reference