Command Reference
R&S
®
ZNB/ZNBT
1019
User Manual 1173.9163.02 ─ 38
Manual operation:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
<Wave>[, <CalPort>,
<SourceType>, <SourcePort>]
Selects the wave quantity and the source for the receiver power calibration, starts the
calibration sweep, and applies the receiver power correction.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored because the port
number is specified in the parameter list.
Setting parameters:
<Wave>
AWAVe | BWAVe | B1 | B2 | B3 | B4
AWAVe
–
calibration of reference waves a
1
, a
2
, .... In manual
control, the reference receiver calibration is included in the
source power calibration of each port.
BWAVe
–
calibration of received wave b
1
, b
2
, .... The port num-
ber
<CalPort>
, the used
<SourceType>
, and the
<SourcePort>
number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ...
–
direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No
additional parameters need to be specified. The source for B1 is
Port 2 and vice versa.
<CalPort>
Calibrated port number
Range:
1 to port number of the analyzer
<SourceType>
PORT | GENerator
PORT - internal source at port
<CalPort>
, to be fed to port
<SourcePort>
using an external through connection. If
<CalPort> = <SourcePort>
, an Open or Short standard is
required.
GENerator - external generator no. <source_no>.
<SourcePort>
Number of the port for the internal source or generator.
Range:
1 to port number of the analyzer/number of external
generators.
SCPI Command Reference