GUI Reference
R&S
®
ZNB/ZNBT
463
User Manual 1173.9163.02 ─ 38
DUT
Balanced
port
Balanced
port
Log.
NWA
port
Physical port c
Logical ports j,k
(single-ended)
Physical port d
Logical port i
(balanced)
Physical port a
Physical port b
Imbalance 1 / Imbalance 2
Imbalance 1
selects the balanced logical port i as the input, the single-ended logical
ports j and k as the output and calculates the imbalance parameter
Imb
jk-i
= –(S
ca
– S
da
)/(S
cb
– S
db
)
Imbalance 2
selects the balanced logical port i as the output, the single-ended logical
ports j and k as the input and calculates the imbalance parameter
Imb
i-jk
= –(S
ac
– S
ad
)/(S
bc
– S
bd
)
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "IMB<i>-<j><k>" |
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "IMB<i>-<j><k>" |
5.9.9 Stability Tab
Selects one of the three two port stability factors K, μ
1
or μ
2
as measured quantities. A
typical application of stability factors is to assess the stability of an amplifier. Stability
factors cannot be calculated in balanced port configurations.
Background information
Refer to
Chapter 4.3.7, "Stability Factors"