Getting Started
R&S
®
ZNB/ZNBT
64
User Manual 1173.9163.02 ─ 38
To set the sweep range use one of the following methods.
Define "Start" and "Stop" values:
1. Create two normal markers, e.g. the markers Mkr 1 (M1) and Mkr 2 (M2), and
place them to the desired start and stop values of the sweep range.
2. Activate "M1" and tap "TRACE > MARKER > Set by Marker > Start = Marker".
3. Activate "M2" and tap "TRACE > MARKER > Set by Marker > Stop = Marker".
Use a definite "Span:"
1. Create a marker and set it to delta mode.
The analyzer automatically creates a reference marker in addition to the delta
marker.
2. Place the reference marker to the desired start value of the sweep range.
3. Set the value of the delta marker equal to the desired (positive or negative) span.
4. Activate the delta marker and tap "TRACE > MARKER > Set by Marker > Span =
Marker".
To move the trace relative to the vertical axis proceed as follows:
1. Create a normal marker, e.g. the marker Mkr 1 (M1), and place it to a particular
trace point. E.g. you can use the "Marker Search" functions to locate a maximum or
minimum on the trace.
2. Select "TRACE > MARKER > Set by Marker > Max = Marker" to move the trace
towards the upper diagram edge, leaving the values of the vertical divisions
("Scale/Div") and the overall vertical scale unchanged. Analogously, select "Min =
Marker" to move the trace towards the lower diagram edge, or select "Ref Val =
Marker" to move the trace towards the "Ref Value".