5.13.2.1
Controls on the Ripple Test Tab
The "Def. Ripple Test..." button opens the "Define Ripple Test" dialog (see
ter 5.13.2.2, "Define Ripple Test Dialog"
Show Ripple Limits
Shows or hides the ripple limit lines associated with the active trace in a Cartesian dia-
gram area. The vertical positions of the ripple lines are re-calculated after each sweep;
only their stimulus range and distance (the ripple limit) is fixed.
Note:
Display of the ripple limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
Ripple Check
Switches the ripple limit check of the active trace on or off.
When the limit check is switched on, a movable info field shows the pass/fail informa-
tion and the measured ripple in each ripple limit range. If the ripple limit check fails at a
measurement point, the point is marked with a colored square. The "Limit Fail Trace"
color is defined in the "Define User Color Scheme" dialog ("Display > Display Config. >
Define User Color..."). An acoustic signal ("Global Beep") and a TTL signal indicating
pass or fail can be generated in addition.