GUI Reference
R&S
®
ZNB/ZNBT
373
User Manual 1173.9163.02 ─ 38
5.5.3 Mode Tab
Optimizes the measurement process.
5.5.3.1
Controls on the Mode Tab
The controls on the "Mode" tab are organized in sections "Driving Mode", "Image
Suppr." and "AGC Mode".
Driving Mode
Determines the order of partial measurements and sweeps.
●
In "Auto" mode, the analyzer optimizes the display update: Fast sweeps are per-
formed in "Alternated" mode, slower sweeps in "Chopped" mode.
●
In "Alternated" mode, a partial measurement is performed at all sweep points (par-
tial sweep) before the hardware settings are changed and the next partial mea-
surement is carried out in an additional sweep. This is usually faster than "Chop-
ped" mode.
●
In "Chopped" sweep mode, the analyzer completes the necessary sequence of
partial measurements at each sweep point and obtains the result (measurement
point) before proceeding to the next sweep point. A trace is obtained from the
beginning of the sweep.
The "Driving Mode" setting is also used during a system error correction. For channels
which require a single partial measurement only, the driving mode settings are equiva-
lent. See also
Chapter 4.1.4.1, "Partial Measurements and Driving Mode"
Remote command:
Channel Config Softtool