Command Reference
R&S
®
ZNB/ZNBT
859
User Manual 1173.9163.02 ─ 38
Example:
:CONTrol1:GPIO1:RANGe 20.0mA
:CONTrol1:GPIO1:SHUNt?
Returns 10 [Ω]
:CONTrol1:GPIO1:RANGe 2mA
:CONTrol:GPIO1:TEST:CLOCk:SHUNt?
Returns 100 [Ω]
:CONTrol1:GPIO1:RANGe 0.2mA
:CONTrol:GPIO1:SHUNt?
Returns 1000 [Ω]
Usage:
Query only
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
2 μA
20 μA
200 μA
2 mA
20 mA
Shunt
100 kΩ
10 kΩ
1 kΩ
100 Ω
10 Ω
CONTrol<Ch>:GPIO<Port>:VOLTage[:DEFault]
<Voltage>
Sets the output voltage of the respective GPIO port.
This voltage is applied using
CONTrol<Ch>:GPIO:VOLTage:OUTPut
Suffix:
<Ch>
.
Channel number
<Port>
GPIO port number
Parameters:
<Voltage>
Range:
-7 to +15 V
Increment: 5 mV
Default unit: V
Manual operation:
See
CONTrol:HANDler[:EXTension]:INDex:STATe
<Boolean>
Selects the digital signal that is routed to pin 20 of the Universal Interface connector.
Parameters:
<Boolean>
ON - /INDEX signal at pin 20
OFF - /PORT_B6 signal at pin 20
*RST:
n/a (default: OFF)
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
SCPI Command Reference