GUI Reference
R&S
®
ZNB/ZNBT
454
User Manual 1173.9163.02 ─ 38
Order
Defines the order of the intermodulation product.
Remote command:
Assign a measurement result to an
existing
trace:
CALCulate<Ch>:PARameter:MEASure
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
Side
Defines the position of the intermodulation product relative to the lower and upper
tones.
"Major"
Denotes the lower or upper intermodulation product, whichever is
larger. The "Major" intermodulation product reveals the worst-case
performance of the DUT.
"Upper"
The "Upper" intermodulation products are measured at frequencies
above the upper tone.
"Lower"
The "Lower" intermodulation products are measured at frequencies
below the lower tone.
Remote command:
Assign a measurement result to an
existing
trace:
CALCulate<Ch>:PARameter:MEASure
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
Relative
If "Relative" is selected, the intermodulation product is displayed in dB units relative to
the measured lower tone level at the DUT output ("Lower Tone at DUT Out"). The rela-
tive result is often termed "intermodulation suppression".
Remote command:
Assign a measurement result to an
existing
trace:
CALCulate<Ch>:PARameter:MEASure
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
Meas Softtool