Command Reference
R&S
®
ZNB/ZNBT
1157
User Manual 1173.9163.02 ─ 38
Parameters:
<Unused>
Currently this parameter is ignored and always returns an empty
string
''
.
<Driver>
Matrix driver name indicating the matrix type (string parameter).
Standard drivers:
The supported drivers are displayed in the
"Driver" dropdown-list of the "Add External Switchmatrix" dialog.
Each list entry corresponds to a matrix driver file (
*.matrix
)
located in the
Resources\ExtDev
subdirectory of the analyz-
er's program directory. Specify the <Driver> as the driver's file
name without the
*.matrix
extension, e.g. as
"ZV-Z81-05"
.
Non-standard driver location:
You may also specify the full
absolute path to an appropriate driver file, e.g.
"C:\Program
Files\Rohde-Schwarz\Vector Network
Analyzer\Resources\ExtDev\ZV-Z81-05.matrix"
.
<Interface>
Interface type (string parameter): either
"USB"
,
"LAN"
or
"DIRECTCTRL"
.
<Address>
Interface-specific address (string parameter).
Example:
:SYST:COMM:RDEV:SMATrix1:DEF '', 'ZN-Z84-42',
'USB', '101142'
Register switch matrix no.1: type R&S ZN-Z84 (2x24) with serial
number
101142
at USB interface.
:SYST:COMM:RDEV:SMATrix2:DEF '', 'ZV-Z81-05',
'LAN', '10.10.10.10'
Register switch matrix no.2: type R&S ZV-Z81 (model .05) at IP
address 10.10.10.10.
:SYST:COMM:RDEV:SMATrix3:
DEF '', 'ZN-Z84-42', 'DIRECTCTRL', '1319.4500K02:
:101143'
Register switch matrix no.3: type R&S ZN-Z84 (2x24) with serial
number
101143
at Direct Control interface.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:MLVNa
{<Matrix
VNA Port Label>, <VNA Port Number>}
Sets/gets the RF connections between switch matrix and VNA. Similar to
COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:MVNA
, but uses the front
panel labels of the matrix VNA ports instead.
By default, these labels are only available for some matrix types.
Note:
a manual RF connection configuration must be performed within a transaction,
enclosed by the commands
SYSTem:COMMunicate:RDEVice:SMATrix:
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
. The modified configuration will not be applied until the end of this transaction.
SCPI Command Reference