Command Reference
R&S
®
ZNB/ZNBT
1082
User Manual 1173.9163.02 ─ 38
Example:
*RST
Reset the instrument, activating a frequency sweep with the S-
parameter S
21
as a measurement result for channel and trace
no. 1.
SWEep:TIME?
Query total sweep time.
SWEep:DWEL 1
SWEep:DWEL:IPOint ALL
Set a delay of 1 s for each partial measurement.
SWE:TIME?
Query total sweep time. The time is extended by the delay times
the total number of sweep points (one partial measurement per
sweep point required).
Manual operation:
See
[SENSe<Ch>:]SWEep:DWELl:IPOint
<InsertionPoints>
Defines whether the measurement delay (previously defined via
) is inserted before all partial measurements or before the first partial
measurement only.
Suffix:
<Ch>
.
Channel number
Parameters:
<InsertionPoints>
ALL | FIRSt
Insertion before all or before the first partial measurement
*RST:
ALL
Example:
See
Manual operation:
See
"All Partial Meas'ments / First Partial Meas'ment"
[SENSe<Ch>:]SWEep:GENeration
<arg0>
Sets/gets the sweep mode for linear frequency sweeps.
For segmented sweeps this can be overridden by segment-specific sweep modes
(
[SENSe<Ch>:]SEGMent<Seg>:SWEep:GENeration
).
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0>
STEPped | ANALog
Stepped mode or swept mode (=ANALog).
*RST:
STEPped
Manual operation:
See
SCPI Command Reference