Power Mode at Sweep End
The analyzer offers three power modes at sweep end:
●
"Auto" results in the shortest measurement time (default setting). If enabled, at
sweep end the output power of the first measurement point is restored. The config-
ured
is not applied.
●
"Reduce" is intended for measurements on sensitive DUTs (primarily: power
sweeps). If enabled, at sweep end the output power of the driving port is reduced
as if the channel base power was set to its minimum possible value. The config-
ured settling delay is applied.
●
"Keep" is intended for power sweeps. If enabled, at sweep end the output power of
the last measurement point is kept. The configured settling delay is applied.
See
Note:
●
Compared to "Auto" mode, "Reduce" and "Keep" may result in significantly longer
measurement times - in particular if an extended settling delay is used.
●
The output power is not altered if there's only a single channel with a single driving
port, performing a Time or CW Mode sweep.
●
In triggered mode the analyzer always uses the settings of the first measurement
point while waiting for the trigger signal.
Remote command:
SOURce<Ch>:POWer<PhyPt>:SWEepend:MODE
Settling Delay / Reset Delay
If
is set to "Reduce" or "Keep" , the "Settling Delay" defines
the time between
for an illus-
tration.
Use the "Reset Delay" button to adjust the "Settling Delay" to its default value.
Remote command:
SOURce<Ch>:POWer<PhyPt>:SWEepend:SDELay
5.19.1.3
Info Dialog
The "Info" dialog displays information about the instrument and its operation. All func-
tions are primarily intended for error diagnostic and service purposes; see
on page 1248. Many "Info" tabs also display softkeys
for printing the contents or saving them to a file.
Access:
SETUP > "Setup" > "Info..."