GUI Reference
R&S
®
ZNB/ZNBT
269
User Manual 1173.9163.02 ─ 38
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select
"Default Port Assignment".
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Prev/Next
Navigates between the port assignments. Only available if multiple port assignments
are required.
Apply/Cancel
Apply the calculated system error correction to the active channel (or to all channels in
the active recall set, if all channels were calibrated).
The Apply button is enabled as soon as calibration sweeps have been successfully
performed for all required port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Extended cal unit settings:
MMEMory:AKAL:FACTory:CONVersion
MMEMory:AKAL:USER:CONVersion
SYSTem:COMMunicate:AKAL:CONNection
SYSTem:COMMunicate:AKAL:MMEMory[:STATe]
5.4.1.3
Calibration Unit Wizard (MultiCal)
guides you through the setup and
simultaneous acquisition of multiple automatic system error corrections.
Access:
CHANNEL – CAL > "Start Cal" > "Calibration" – "Start... (Cal Unit)"
The wizard proceeds through the following steps:
1.
"Ports":
Define the system error corrections to be performed (test ports, calibra-
tion unit, characterization, calibration type)
Cal Softtool