Command Reference
R&S
®
ZNB/ZNBT
992
User Manual 1173.9163.02 ─ 38
Note:
The power calibration must be performed "after" the system error correction; see
example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<TestPort>
Test port number. The port number must match the port number
selected via
[SENSe:]CORRection:COLLect:PMETer:ID
;
see example.
Example:
Suppose that a power meter no. 1 is configured and USB-con-
nected to the analyzer.
SENSe:CORRection:COLLect:PMETer:ID 1, 2
Select power meter no. 1, to be connected to Port 2, for the
SMARTerCal.
--> Connect the calibration unit between ports 1 and 2 of the
analyzer,
SENSe:CORRection:COLLect:AUTO:PORTs:TYPE
PFNPort, '', 1, 1, 2, 2
Perform an automatic full 2-port (UOSM) calibration at the ana-
lyzer test ports 1 and 2 using the calibration unit's default cali-
bration kit file and default port assignments (ports 1 and 2 of the
cal unit).
--> Remove the cal unit and connect the power meter to Port 2
of the analyzer.
SENSe:CORRection:COLLect:AUTO:POWer 2
Perform an additional receiver power calibration at Port 2. This
completes the
Usage:
Setting only
Manual operation:
See
"Start Cal Sweep / Abort Sweep"
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
<CalType>[, <Characterization>,
<TestPort1>, <TestPort2>...]
Selects and initiates an automatic calibration at arbitrary analyzer and cal unit ports.
This command also selects the calibration type. A progress monitor for the calibration
sweeps is displayed.
Tip:
If the test setup contains a high attenuation the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
.
If several calibration units are connected, use
to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
SCPI Command Reference