Command Reference
R&S
®
ZNB/ZNBT
987
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Type>
THRough | OPEN | SHORt | OSHort | OSHort1 | OSHort2 |
OSHort3 | MATCh | NET | ATT | REFL | SLIDe | ISOLation |
LINE | LINe1 | LINe2 | LINe3 | UTHRough | POWer
Standard types: Through, Open, Short, Match, Symmetric Net-
work (NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe),
Line1 (LINE1 and LINE are synonymous), Line2 and Line3 (esp.
for TRL calibration), Offset Short 1 to 3 (OSHort), Unknown
Through, power meter (for SMARTerCal).
ISOLation is not a physical standard: to measure the isolation
(supported for transmission normalization and TOSM only!) it is
recommended to suitably terminate the related test ports (e.g.
with 50 Ω loads).
<TestPort>
Test port number. For a transmission standard (through, line,
attenuation, symmetric network) or an adapter used as a
"through" the input and output port numbers must be specified.
For reflection standards, only one port number is required.
*RST:
n/a
<SecondPortOrAdapter>
For a transmission standard or an adapter used as a "through"
this parameter specifies the second test port.
For reflection measurements with an adapter connected
between port and standard, set it to ON.
Otherwise set it to OFF or simply omit it.
<Dispersion>
Optional status parameter for UTHRough standard:
OFF - unknown through standard is non-dispersive.
ON - unknown through standard is dispersive.
*RST:
OFF
SCPI Command Reference