GUI Reference
R&S
®
ZNB/ZNBT
316
User Manual 1173.9163.02 ─ 38
Add
Adds a new SMARTerCal.
The "Add" button opens the
ddialog without pre-selected
ports.
Delete
Deletes the selected SMARTerCal from the list of configured SMARTerCals.
Modify
Edits the selected SMARTerCal: opens the
dialog with the
corresponding ports and calibration type pre-selected.
Next
Step 2: Connectors and Cal Kits
Inactive as long as no calibration is defined.
Define Calibration (Dialog)
Selects the calibrated analyzer ports, the calibration type and specifies the power
meter connection.
Background information
Refer to
Chapter 4.5.7.2, "Calibration Types"
Access:
CHANNEL > CAL > Start Cal > SMARTerCal
–
Start... (Manual) > Add/Modify
The layout of the main panel depends on the number of test ports available.
Figure 5-23: Define SMARTerCal Dialog (>4 ports)
Ports
Selects the test port(s) to be calibrated.
It is possible to select any combination of two or more test ports. If you're only interes-
ted in a single port
p
, perform a two-port SMARTerCal for a port pair including
p
(as
Power Port).
Cal Softtool