GUI Reference
R&S
®
ZNB/ZNBT
311
User Manual 1173.9163.02 ─ 38
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete
Cal Kit
Selects a cal kit for the connector at each selected physical port. The drop-down list
contains all available calibration kits for the selected connector type.
The assignment of a calibration kit to a connector type must be the same for all physi-
cal ports: If a calibration kit is changed, the R&S
ZNB/ZNBT automatically assigns the
new kit to all ports with the same connector type.
Use "Import Cal Kit..." to add new kits to the list.
Remote command:
[SENSe:]CORRection:CKIT:SELect
Same Connector All Ports / Same Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some
multi-port calibration types, the port connector types must be equal, e.g. because they
require a Through standard with known characteristics.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Kit" dialog that allows to import a cal kit file. For back-
ground information see
Chapter 4.5.2.4, "Cal Kit Files"
Remote command:
Back
Go back to
Step 2: Connectors and Cal Kits
.
Start
Start
Step 3: SMARTerCal
Allows to acquire error correction data for every required port (pair) and calibration
standard (where "required" depends on the selected ports and calibration type) as well
as power correction data from the power meter. The order of the system error correc-
tion and power calibration sweeps is arbitrary; ensure that you always connect the
proper equipment.
On "Apply" the R&S
ZNB/ZNBT calculates the system error correction (error terms)
from the measurement data of the standards and uses the power meter result to cor-
rect the absolute receiver powers and applies the SMARTerCal results to the active
channel.
Access:
CHANNEL > CAL > Start Cal > SMARTerCal
–
Start... (Manual) > Next >
Next
Cal Softtool