Command Reference
R&S
®
ZNB/ZNBT
1007
User Manual 1173.9163.02 ─ 38
Example:
CORRection:CONNection:CATalog?
Query connector types. Possible response:
'N 50 Ohm,N 75
Ohm,7 mm,3.5 mm,2.92 mm,2.4 mm,1.85
mm,7-16,Type F (75),BNC 50 Ohm,BNC 75
Ohm,SMA,4.3-10,User Conn 1'
CORRection:CKIT:CATalog? 'N 50 Ohm'
Query cal kits for N (50
Ω
) connector types. Possible response:
'N 50 Ohm Ideal Kit,3653,85054D,ZV-Z121,ZCAN 50
Ohm,ZV-Z21 typical,85032B/E,85032F,85054B,New
Kit 1'
. We assume that a cal kit
New Kit 1
with label
2012-05-25
was created before.
CORRection:CKIT:LCATalog? 'N 50 Ohm'
Query cal kits for N (50
Ω
) connector types with their labels.
Possible response:
'N 50 Ohm Ideal
Kit,,3653,,85054D,,ZV-Z121,,ZCAN 50 Ohm,,ZV-Z21
typical,,85032B/E,,85032F,,85054B,,New Kit
1,2012-05-25'
. A sequence of two commas means that the
preceding cal kit has no label assigned.
CORRection:CKIT:STANdard:CATalog? '85032B/E'
Query standards in cal kit named 85032B/E. Possible response:
'MOP,FOP,MSH,FSH,MMMT,FMMT,MFTH'
CORRection:CKIT:STANdard:LCATalog? 'New Kit
1','2012-05-25'
Query standards in user cal kit named
New Kit 1
labelled
2012-05-25
. Possible response:
'MOP,FOP,MSH(P1),FSH,MMMT,FMMT,MSM,FSM,MREF,FREF,
MOSHORT2,FOSHORT2,MOSHORT3,FOSHORT3,MMTH,MMTH(P2P3)
...'
(P1 etc. denote restricted port assignments).
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:CONNection:DELete
<ConnectorName>
Deletes a user-defined connector type named
<ConnectorName>
.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<ConnectorName>
Name of the user-defined connectors, string parameter.
Example:
See
[SENSe<Ch>:]CORRection:CONNection
Usage:
Setting only
Manual operation:
See