GUI Reference
R&S
®
ZNB/ZNBT
252
User Manual 1173.9163.02 ─ 38
Remote command:
MMEMory:LOAD:SEGMent
MMEMory:STORe:SEGMent
Displayed Columns Dialog
The "Displayed Columns" dialog selects the segment-specific channel settings that the
analyzer displays in the
and its "Point List" (opened via
)
Access:
All settings can be adjusted in the "Define Segments" dialog. By default, the first sweep
segment is created with the channel settings defined for unsegmented sweep types.
When any further sweep segment created, it uses the channel settings of the previ-
ously active segment.
Related information
Refer to the following sections:
●
Chapter 5.3.2.2, "Define Segments Dialog"
●
Table Columns
Each selected (checked) option adds a column to the segment list.
●
"Name" adds a column to assign a name to each segment. A segment name is a
string that may contain letters, numbers and special characters.
●
"Power (Pb)" defines the internal source power (channel base power) for each indi-
vidual sweep segment. See
●
"Meas Bandwidth" defines the IF filter bandwidth for each individual sweep seg-
ment. See
●
"Frequency Sweep Mode" allows to define whether the "Stepped" or "Swept" mode
shall be used for individual sweep segments (see
Swept mode is only supported for linear frequency sweeps. Even for linear fre-
quency sweeps there are measurement setups that are incompatible with swept
mode: an information popup displays the corresponding error code and description
(see
[SENSe<Ch>:]SWEep:GENeration:ANALog:CONDition?
).