Concepts and Features
R&S
®
ZNB/ZNBT
81
User Manual 1173.9163.02 ─ 38
●
, in order to reduce "crosstalk" between
the DUTs a frequency offset can be applied between the corresponding port
groups (see
"Parallel Measurements with Frequency Offset"
●
Parallel
Frequency Conversion Measurements
require identical conversion settings
for all port groups
●
Parallel
not
supported
●
Parallel measurements are
not
Arbitrary Drive Port Order (R&S
ZNBT only)
When performing parallel measurements on multiple DUTs, by default the R&S
ZNBT
firmware automatically selects the ports that shall driven in parallel. However, due to
possible crosstalk between DUT ports, it may be beneficial to prevent some ports from
being driven in parallel. This could be achieved by changing the cabling of the DUT,
but this may not always be possible - for example if the same cabling shall be used for
different measurements and different kinds of DUTs. Furthermore changing the cabling
requires a recalibration of the affected channels.
In order cope with this problem the R&S
ZNBT firmware allows to restrict or even com-
pletely define these "drive port sets" by specifying a drive port order for selected DUTs.
Basic Concepts