GUI Reference
R&S
®
ZNB/ZNBT
337
User Manual 1173.9163.02 ─ 38
Remote command:
Two Port Config...
Opens the "Two Port Configuration" dialog to define transmission coefficients for the
"Two Port at DUT" and the "Two Port at Power Meter" test setups. The button is dis-
abled if "No Coefficients" is active. See
"Two Port Configuration Dialog"
Two Port Configuration Dialog
The "Two Port Configuration" dialog defines the transmission characteristics of an
additional two-port in the calibrated frequency range.
Access:
Power Meter Transmission Coefficients Dialog
> "Two Port Config..."
Frequency / Transm. Coefficients: Insert, Delete, Delete All
The required two-port information is a list of transmission coefficients at different fre-
quency values (power loss list). The buttons in the dialog provide different ways of cre-
ating and modifying the list. Use "Insert", "Delete", "Delete All" to edit the list manually.
In a power, time or CW mode sweep, one point at the fixed CW frequency is sufficient.
In a frequency sweep it is possible to enter several coefficients to account for a fre-
quency-dependent attenuation. Transmission coefficients are interpolated between the
frequency points and extrapolated, if necessary. If no transmission coefficient is
defined at all, the R&S
ZNB/ZNBT assumes a 0 dB attenuation across the entire fre-
quency range, which is equivalent to an ideal through connection or selecting "No
Coefficients" in the
Power Meter Transmission Coefficients Dialog
.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:INSert<ListNo>
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DEFine<ListNo>
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:COUNt?
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>[:
DUMMy]
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>:
ALL
Cal Softtool