![Rohde & Schwarz ZNB20 User Manual Download Page 973](http://html.mh-extra.com/html/rohde-and-schwarz/znb20/znb20_user-manual_1477889973.webp)
Command Reference
R&S
®
ZNB/ZNBT
958
User Manual 1173.9163.02 ─ 38
Example:
See
7.3.15 [SENSe:] Commands
The
[SENSe:]...
commands affect the receiver settings of the R&S
ZNB/ZNBT.
7.3.15.1
[SENSe:]AVERage...
The
[SENSe:]AVERage...
commands set sweep averaging parameters. The sweep
average is a noise-reduction technique which consists of calculating each measure-
ment point as an average of the same measurement point over several consecutive
sweeps.
In contrast to the sweep count (for single sweep mode,
), averaging is always channel-specific. Both features are completely independent from
each other.
.................................................................................... 958
.......................................................................................958
..................................................................................... 959
...................................................................................... 959
[SENSe<Ch>:]AVERage[:STATe]
<Boolean>
Enable or disable the sweep average.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON | OFF - enables or disables the automatic calculation of the
sweep average over the specified number of sweeps
(
*RST:
ON
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:CLEar
Starts a new average cycle, clearing all previous results and thus eliminating their
effect on the new cycle.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference