Concepts and Features
R&S
®
ZNB/ZNBT
160
User Manual 1173.9163.02 ─ 38
Calibration unit
Frequency range
Connector type
No. of ports
Order no.
R&S ZN-Z152
100 kHz to 8.5 GHz
SMA (f)
6
1319.6003.36
R&S ZN-Z153
100 kHz to 8.5 GHz
SMA (f)
4
1319.6178.34
R&S ZN-Z154
100 kHz to 8.5 GHz
SMA (f)
6, 12, 18 or 24
1319.5120.02
Options R&S
ZN-Z154-B22, -B32 and -B42 (order nos 1319.5136.22, 1319.5136.32 and 1319.5136.42)
extend the R&S
ZN-Z154 base unit with additional ports 7-12, 13-18 and 19-24, respectively.
R&S ZV-Z51
300 kHz to 8 GHz
3.5 mm (f)
4
1164.0515.30
R&S ZV-Z51
300 kHz to 8 GHz
type N (f)
4
1164.0515.70
R&S ZV-Z52
10 MHz to 24 GHz
3.5 mm (f)
4
1164.0521.30
R&S ZV-Z52
100 kHz to 18 GHz
type N (f)
4
1164.0521.70
R&S ZV-Z53
300 kHz to 18 GHz
type N (f)
2
1164.0473.72
R&S ZV-Z53
300 kHz to 24 GHz
3.5 mm (f)
2
1164.0473.32
R&S ZV-Z54
10 MHz to 40 GHz
2.92 mm (f)
2
1164.0467.92
R&S ZV-Z55
10 MHz to 50 GHz
2.4 mm (f)
2
1164.0480.42
R&S ZV-Z58
300 kHz to 8 GHz
type N (f)
8
1164.0638.78
R&S ZV-Z58
300 kHz to 8 GHz
3.5 mm (f)
8
1164.0638.38
R&S ZV-Z58
10 MHz to 20 GHz
3.5 mm (f)
6
1164.0450.36
R&S ZV-Z59
10 MHz to 20 GHz
3.5 mm (f)
6
1164.0450.36
The units contain calibration standards that are electronically switched when a calibra-
tion is performed. The calibration kit data for the internal standards are also stored in
the calibration unit, so that the analyzer can calculate the error terms and apply the cal-
ibration without any further input.
Advantages of automatic calibration
Automatic calibration is generally faster and more secure than manual calibration,
because:
●
There is no need to connect several standards manually. The number of connec-
tions to be performed quickly increases with the number of ports.
●
Invalid calibrations due to operator errors (e.g. wrong standards or improper con-
nections) are almost excluded.
●
No need to handle calibration kit data.
●
The internal standards don't wear out because they are switched electronically.
Limitations of automatic calibration
Calibration types TOM, TSM, TRM, TRL, TNA are not available.
Calibration