GUI Reference
R&S
®
ZNB/ZNBT
287
User Manual 1173.9163.02 ─ 38
The upper part of the panel shows the previously configured calibration(s). Select the
appropriate one to get access to the related ports. For the individual calibrations the
functionality is the same as described for the "SingleCal" version of the wizard (see
"Step 2: Connectors and Cal Kits"
Back
Go back to
.
Start
Start
.
Step 3: Calibration
Allows to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration types.
Same logic as in the single calibration version of the wizard (see
The possibility to "share" measurements between the calibrations and the "Reduced
Through" logic help to keep the number of measurements as small as possible (see
Chapter 4.5.1.11, "Full n-Port Calibration with Reduced Number of Through Connec-
tions"
5.4.1.6
Power Cal Wizard
The "Power Cal" wizard displays a single screen showing the current source ("Power")
and receive ("Meas. Receiver") ports. It allows you to perform scalar source power cali-
brations (flatness calibrations) or measurement receiver calibrations, based on the cur-
rent power calibration settings (see
Chapter 5.4.3, "Pwr Cal Settings Tab"
Cal Softtool