![Rohde & Schwarz ZNB20 User Manual Download Page 411](http://html.mh-extra.com/html/rohde-and-schwarz/znb20/znb20_user-manual_1477889411.webp)
GUI Reference
R&S
®
ZNB/ZNBT
396
User Manual 1173.9163.02 ─ 38
OK
Activates the scalar mixer mode and closes the dialog. A "Mix" label in the channel list
indicates that a mixer measurement is active.
disables the mixer measurement.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion
5.5.6.4
SMARTerCal (Cal Unit) Wizard for Mixer Mode
Allows to perfom a variant of an automatic "SMARTerCal" that is specially tailored to
the "Mixer Mode".
Access:
CHANNEL – CHANNEL CONFIG > "Mixer Mode" > "SMARTerCal" – "Start...
(Cal Unit)"
Offers similar functionality as the general
lowing differences:
●
Always all channels are calibrated
●
The affected test ports are already preselected and labeled according to their role
in the mixer measurement
●
The adequate calibration type is already preselected
●
A source flatness calibration can be automatically included as a the final calibration
step
Background information
Refer to
on page 172 for background information
The wizard proceeds as follows:
1. Specify the remaining (non-fixed) properties of the automatic SMARTerCal
(required hardware, ports, connections etc.)
Channel Config Softtool