Marker Name
Assigns a (new) name to the active marker. Marker names may exceed the length of
the input box and contain letters, numbers, blanks and special characters.
Remote command:
CALCulate<Chn>:MARKer<Mk>:NAME
CALCulate<Chn>:MARKer<Mk>:REFerence:NAME
Marker Format
Defines the formatting of the active marker in the movable marker info field.
Note that in contrast to previous releases "Default" now means the marker is formatted
according the related trace's
, which may not be set to "Default"
(i.e. formatted according to the selected trace format).
For background information on marker formats see
, if the transformation between trace format and marker for-
mat requires a concrete stimulus value, some result values in the marker info field may
be unavailable. Those values are displayed as a sequence of dashes (------).
Remote command:
CALCulate<Chn>:MARKer<Mk>:FORMat
Marker Style
Defines how the selected marker is displayed on the screen.
Remote command:
n/a
Discrete
Discrete mode means that a marker can be set to discrete sweep points only. If dis-
crete mode is switched off, the marker can be positioned on any point of the trace, and
its response values are obtained by interpolation.