GUI Reference
R&S
®
ZNB/ZNBT
514
User Manual 1173.9163.02 ─ 38
Checks and Messages in the Dialog
After each port or channel selection, the R&S
ZNB/ZNBT checks the channel data for
compatibility with the trace export conditions. If data from "All Channels" are exported,
every single channel must contain a compatible set of traces; see
The "OK" button is available only if no error message is displayed in the dialog.
Select / Select All / Deselect All
Selects the ports to be considered for the S-parameter export.
Example:
With ports 1 and 2 selected, S-parameters S
11
, S
12
, S
21
and S
22
will be
exported.
Remote command:
Assign
Selects the port number assignment in the created
*.s<n>p
file. By default, analyzer
and
*.s<n>p
port numbers are identical. You can interchange the port assignment in
order to change the order of the S-parameters in the created ´Touchstone file. Each of
the analyzer port numbers must be assigned to one
*.s<n>p
port number.
Active Channel / All Channels
Selects data export for the active channel or for all channels.
Reset Port Assignments
Restores the identity between original and assigned port number.
5.13 Lines Softtool
The "Lines" softtool allows to define limits for the measurement results, visualize them
in the diagrams and activate/deactivate the limit check. The analyzer provides upper,
lower, ripple and circle limits. In addition the "Lines" softtool provides functions to limit
Lines Softtool