Concepts and Features
R&S
®
ZNB/ZNBT
83
User Manual 1173.9163.02 ─ 38
●
Please note that in parallel measurement with frequency offset the firmware uses a
modified IF as compared to measurements not using this mode. Because this
modified IF requires a special calibration it is essential to perform the
with the same Frequency Offset settings as for the actual measurement; otherwise
the calibration is turned off (see
●
For a R&S
ZNBT parallel measurement with frequency offset is only possible if the
configured DUTs use different sources. Hence the R&S
ZNBT must be equipped
with more than 8 ports and one DUT must be a subset of ports 1-8, the other a
subset of ports 9-24.
4.1.4.3
Stimulus and Sweep Types
The function of the STIMULUS hardkeys START, STOP, CENTER and SPAN depends
on the sweep type.
Table 4-1: Function of STIMULUS keys
Sweep type
START (unit)
STOP (unit)
CENTER (unit)
SPAN (unit)
"Lin Freq"
"Start Frequency" (Hz)
"Stop Frequency" (Hz)
"Center Frequency"
(Hz)
"Span Frequency" (Hz)
"Log Freq"
"Start Frequency" (Hz)
"Stop Frequency" (Hz)
–
–
"Segmented"
–
–
–
–
"Power"
"Start Power" (dBm)
"Stop Power" (dBm)
"CW Frequency" (Hz)
"CW Frequency" (Hz)
"CW Mode"
"CW Frequency" (Hz)
"Number of Points" (-)
"CW Frequency" (Hz)
"CW Frequency" (Hz)
"Time"
"CW Frequency" (Hz)
"Stop Time" (s)
"CW Frequency" (Hz)
"CW Frequency" (Hz)
The ranges of numerical values must be compatible with the instrument model. The
conditions for the stimulus range depend on the sweep type:
●
"Lin Freq" / "Log Freq" / "Segmented"
The supported frequency range depends on the instrument type; for an overview
refer to
.
The stop frequency must be greater than the start frequency; the span must be
≥
1
Hz. If a stop frequency smaller than the current start frequency is set, then the start
frequency is adjusted and vice versa. A sweep must contain at least two different
sweep points.
●
"Power"
Start and stop power are both entered in absolute units (dBm). Start and stop
power must be different; the stop power must be larger than the start power. If a
stop power smaller than the start power is set, then the start power is adjusted
automatically and vice versa.
The power corresponds to the actual source power at the test ports (channel base
power P
b
). After a port power calibration, this source power is available at the cali-
brated reference plane.
●
"CW Mode"
Basic Concepts