GUI Reference
R&S
®
ZNB/ZNBT
529
User Manual 1173.9163.02 ─ 38
Remote command:
CALCulate<Chn>:LIMit:CIRCle[:STATe]
CALCulate<Chn>:LIMit:CIRCle:FAIL?
CALCulate:LIMit:CIRCle:FAIL:ALL?
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:CLEar
Draw Circle
Activates touchscreen or mouse operation; tap the diagram at one border of the limit
circle and draw the circle to the required size and position.
Remote command:
n/a
Radius / Center X / Center Y
Defines the limit circle by its radius and its center on the X-axis and Y-axis.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:DATA
Global Check
See
TTL 1 Pass / TTL 2 Pass
See
5.13.4 Display Circle Tab
The "Display Circle" functionality allows to limit results displayed in the complex plane
(Smith, Polar) to a user-defined circle.
The controls on this tab are active only if the active trace is currently displayed in com-
plex format.
Lines Softtool