Command Reference
R&S
®
ZNB/ZNBT
1106
User Manual 1173.9163.02 ─ 38
Parameters:
<arg0>
The multiplication factor to calculate the minimum frequency off-
set.
To get the resulting frequency offset, the minimum frequency off-
set is rounded to a multiple of the current frequency step size.
Range:
1 to 100
*RST:
10
Example:
See
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:
Manual operation:
See
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition
This command is used to query the operation mode of Simultaneous Measurement
with Frequency Offset. The list below contains all possible return values of this query.
Please note that this query will return an error if the channel specified by <Ch> does
not contain any trace. This is due to the fact that without traces no measurements are
performed so that no information about Simultaneous Measurement with Frequency
Offset is available.
This query waits until all previous commands are fully executed because these com-
mands may affect the operation mode of Simultaneous Measurement with Frequency
Offset.
Suffix:
<Ch>
.
Channel number.
If unspecified, this is set to 1.
<Grp>
SCPI Command Reference