MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
<TouchstoneFile>[, <Interchange>]
Loads data from a specified two-port (
*.s2p
) Touchstone file defining a single ended
circuit model for deembedding.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA
to load circuit data from the remote client.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Setting parameters:
<TouchstoneFile>
String parameter to specify the name and directory of the loaded
Touchstone file.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
<Interchange>
FPORts | IPORts
Interchange port numbers of loaded
*.s2p
file:
FPORts
(or omitted): the port numbers are not interchanged
IPORts
: interchange port numbers
The parameter must
not
be used for 4-port files.
Example:
CALC:TRAN:VNET:SEND:DEEM:TND FIMP
Select the "Serial .s2p" data circuit model.
MMEM:LOAD:VNET:SEND:DEEM2 'C:\Users\Public
\Documents\Rohde-Schwarz\Vna\VNET\Test.s2p'
Load a Touchstone file and assign it to the physical port no. 2.
Usage:
Setting only
Manual operation:
See
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt>
<TouchstoneFile>[, <Interchange>]
Loads data from a specified two-port (
*.s2p
) Touchstone file defining a single ended
circuit model for embedding.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA
to load circuit data from the remote client.
SCPI Command Reference