Command Reference
R&S
®
ZNB/ZNBT
835
User Manual 1173.9163.02 ─ 38
Parameters for setting and query:
<CircuitModel>
CSL | LSC | CSC | SLC | SCL | SCC | SHLC | GSL | LSG |
GSG | SLG | SGL | SGG
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C cir-
cuit model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA
<Interchange>, <arg1>
Defines an embedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
cuit data from a Touchstone file located at the R&S
ZNB/ZNBT's file system.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Setting parameters:
<Interchange>
FPORts | IPORts
FPORts - the port numbers are not interchanged.
IPORts - Interchange port numbers of loaded trace data.
<arg1>
<block_data>
Content of a Touchstone file (
*.s2p
) in IEEE488.2
.
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G in the "shunt L, shunt C" circuit model for single
ended port deembedding.
In the query form, the <Conductance> parameter must be omitted.
SCPI Command Reference