GUI Reference
R&S
®
ZNB/ZNBT
330
User Manual 1173.9163.02 ─ 38
Test Port Assignment
Defines the assignment between test ports and cal unit ports. In the default "Manual"
assignment, VNA ports and cal unit port numbers match. If you decide to use a differ-
ent assignment, you can auto-detect the actual assignment ("Automatic") or select the
analyzer port numbers manually. Auto-detection may fail e.g. because of a high attenu-
ation in the signal path.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs
Take OSM / Take All OSM and Through
Starts a calibration sweep for the selected port(s). "Take All OSM and Through" ini-
tiates a series of calibration sweeps; the R&S
ZNB/ZNBT acquires a full set of one-port
and two-port data. This is required for the transmission normalizations and for a "One
Path Two Ports" calibration; see
"Dependency between calibration types and charac-
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Next
Step 2: Save Characterization Data
. Next is available as soon as the
R&S
ZNB/ZNBT has acquired characterization data for a single port.
Cal Softtool