Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:DEFine
<StartFreq>, <StopFreq>, <Points>,
<Power>, <SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[,
<LO>, <Selectivity>, <FreqSweepMode>]
Creates or re-defines a sweep segment no.
<Seg>
with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provi-
ded. All settings except
<LO>
can be changed for existing segments using other com-
mands of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
to create a segment with default
channel settings. Use
[SENSe<Ch>:]SEGMent<Seg>:INSert
new segment into the current segment list.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If segment number
<Seg>
already exists, it is replaced by the
new segment.
Parameters:
<StartFreq>
Start frequency of the segment; see
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
Default unit: Hz
<Points>
Number of sweep points in the segment. See
<Power>
Internal source power in the segment. See
Default unit: dBm
<SegmentTime>
Duration of the sweep in the segment. See
DWELl
, this parameter is
replaced by
<MeasDelay>
.
Range:
Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
SCPI Command Reference