Command Reference
R&S
®
ZNB/ZNBT
1067
User Manual 1173.9163.02 ─ 38
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer?
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?
These commands return the center frequency and the span (width) of sweep segment
no.
<Seg>
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:CENT? SPAN?
Query the center frequency and span of the created segment.
The response is
1250000;500000
.
Usage:
Query only
Manual operation:
See
Note:
The frequency range of the sweep segment can be changed via
and
.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt
<StartFreq>
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP
<StopFreq>
These commands define the start and stop frequency of sweep segment no.
<Seg>
.
The sweep segments must be within the frequency range of the R&S
ZNB/ZNBT; see
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<StopFreq>
Start or stop frequency of the sweep.
Increment: 0.1 kHz
Default unit: Hz
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:STAR?; STOP?
Query the start and stop frequency of the created segment. The
response is
1000000;1500000
.
Manual operation:
See
SCPI Command Reference