Command Reference
R&S
®
ZNB/ZNBT
983
User Manual 1173.9163.02 ─ 38
Example:
CORR:COLL:AUTO:PORT '', 1, 2, 2, 4, 4, 1
Perform an automatic 3-port calibration at the analyzer ports 1,
2, and 4 using the calibration unit's default calibration kit file and
ports 2, 4, and 1 of the cal unit.
CORR:COLL:AUTO:PORT:CONN?
Query the actual port assignment. If the cal unit is properly con-
nected according to the previous command, the response is
1,2,2,4,3,0,4,1. A zero means that the corresponding analyzer
port is not connected to any port of the calibration unit.
Usage:
Query only
Manual operation:
See
[SENSe:]CORRection:COLLect:AVERage
<Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see
) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average>
AUTO | MANual
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
<Boolean>
Toggles the
M
ultiple
C
alibration
TYpes
mode that allows calibrate a subset of the
available channels using channel-specific ports and calibration types (see
ter 4.5.8, "Parallel Calibration of Multiple Channels"
SCPI Command Reference