Command Reference
R&S
®
ZNB/ZNBT
785
User Manual 1173.9163.02 ─ 38
Parameters:
<Stop>
Start or stop value of the evaluation range.
Default unit: NN
Example:
See
CALCulate<Chn>:STATistics:DOMain:USER
Manual operation:
See
CALCulate<Chn>:STATistics:EPDelay[:STATe]
<Boolean>
CALCulate<Chn>:STATistics:MMPTpeak[:STATe]
<Boolean>
CALCulate<Chn>:STATistics:MSTDdev[:STATe]
<Boolean>
These commands display or hide the "Phase/El Length" results, the "Min/Max/Peak-
Peak" results, and the "Mean/Std Dev" results in the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - statistical info field on or off.
*RST:
OFF
Example:
See
CALCulate<Chn>:STATistics[:STATe]
Manual operation:
See
"Min/Max/Peak-Peak, Mean/Std Dev/RMS"
CALCulate<Chn>:STATistics:FORMat
<Format>
For complex-valued traces (Smith, Polar) this determines how the MEAN, STDDev,
MAX, MIN, RMS and PTPeak statistics are calculated, see
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Format>
ZVAB | IMPedance | ADMittance
ZVAB
The results are based on unformatted wave quantities (voltages)
IMPedance
The results are based on resistance values
ADMittance
The results are based on conductance vaules
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel
<Stop>
Defines the compression value x for the compression point measurement
(
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
).